Model 5100 Download printable brochures: Model 5100 Oxygen in Process Gas, Model 5100 Carbon Dioxide, Model 5100 Moisture and Methane in Natural Gas, Model 5100 Moisture in Natural Gas, Model 5100 H2O in Gas Streams, Model-5100-CO2-Atex, Model
5100/5100HD H2S in Process Gas, Model 5100 5100HD H2O Measurement, Model 5100 General Moisture
The Model 5100 hosts a single laser platform to measure a single gas analyte. The Model 5100 can measure any one of the following gas analytes: water vapor, methane, hydrogen sulfide, carbon dioxide, and oxygen in process or flue gas. If water vapor is the measured gas analyte, methane can be monitored additionally at percent levels using the same laser.
The Model 5100 is designed to handle sample gases with dew points as high as 55ºC. For higher temperature dewpoint applications the Model 5100HD is recommended.
A key feature of the Model 5100 is the use of a sealed reference cell, which contains a known amount of analyte gas measured, for referencing the emission wavelength of the laser. The use of a reference cell enables both the ability to line-lock the lasers and to continuously verify the instrument calibration.
Model 5100 offers high specificity, selectivity, speed of response and great reliability.
Features and Benefits
-
Noncontact measurement for low maintenance
-
All digital signal processing
-
Web-based interface
-
Connectivity by modbus, ethernet and analog
-
Real-time performance monitoring
-
Fully integrated sample handling
-
Resistant to contamination
-
Reference Cell enables Line Lock and Span Verification
|
|
|
Model 5100 Moisture in Natural Gas Specifications
|
Other specifications available on request.
|
|
Laser Specifications:
|
Class 1m
|
|
Typical Operating Range:
|
0.25 - 60 lb/MMscf / 4 - 1900 mg/m3 (5 to 2500 ppmv)
|
|
Limit of Detection:
|
0.25 lb/MMscf / 4 mg/m3 (5 ppmv)
|
|
Accuracy:
|
2% of reading or ±4 ppmv whichever is greater
|
|
Repeatability:
|
2% of reading or ±4 ppmv whichever is greater
|
|
Ambient Temperature:
|
-20ºC to +50ºC (-4ºF to 122ºF)
|
|
Relative Humidity:
|
0% to 90%, noncondensing
|
|
Sample Flow Rate:
|
1 to 10 SLPM recommended (2 - 20 SCFH)
|
|
Sample Cell Pressure:
|
70 to 170 kPa absolute (10-25 psia)
|
|
Speed of Response:
|
<1 second photometric response. Total system response is dependent on sample flow rate.
|
|
Outputs:
|
4-line x 20-character alphanumeric VF display
Fast Ethernet (IEEE802.3)
RS-485 serial port, isolated (supports Modicon Modbus RTU)
(1) isolated 4-20 mA loop-powered analog output
(4) dry relay contacts. Contact rating 30 VAC, 60 VDC, 100 VA resistive
|
|
Electrical Requirements:
|
120 VAC (108-132V); 47-63 Hz, or
240 VAC (216-264), 47-63 Hz
|
|
Power Requirements:
|
<25W;with optional heater 105W
|
|
Physical (HxWxD):
|
86.88 cm x 43.42 cm x 21.17 cm (34.2 in x 17.1 in x 8.34 in)
|
|
Weight:
|
25 kg (55 lb)
|
|
Enclosure:
|
IP65, NEMA 4
|
|