AMETEK offers on-line process analyzers for the accurate determination of the trace moisture content of gases. Analysis ranges and accuracy vary by model from as high as 2500 parts per million by volume down to <1 part per billion by volume. In addition, AMETEK provides a line of products based on tunable diode laser technology.
|
|
Model 5920 UHP Moisture Analyzer Series Download printable brochure.
Building upon the success of our quartz-crystal moisture measurement technology, AMETEK’s Model 5920 is an easy-to-use process moisture analyzer that offers a truly remarkable combination of performance features: multi-gas compatibility, exceptional accuracy, on-line verification, fast response speed, and wide
measurement range. The 5920 is ideal for critical moisture measurement applications including high purity gas production and semiconductor gases. Equipped with an on-line verification system, this analyzer is designed to rapidly build and maintain operator confidence in its analyses. The verification system allows you to challenge the analyzer’s sensor at will using a zero gas to check its baseline stability to confirm the sensor’s stability and
sensitivity to low ppbv moisture concentrations.
|
|
|
Model 5920UHP Specifications
|
Compatible Gases:
|
Inerts (He, Ar, Ne, Xe, Kr) O2, H2, N2. Contact AMETEK to confirm compatibility with other gases.
|
|
Range:
|
Calibrated from 0 to 150 ppbv. Trend indication to 1000 ppbv.
|
|
Limit of Detection:
|
1.0 ppbv nominal
|
|
Accuracy:
|
±1 ppbv or ±10% of the reading, whichever is greater
|
|
Sensitivity:
|
Better than 1 ppbv
|
|
Response Time:
|
Typically 80% of a 25 ppbv step change in either direction in 10 minutes or less
|
|
Inlet Pressure:
|
138 to 345 kPa (20 to 50 psig). Specified performance is obtained when the inlet gas pressure is maintained within ±17 kPa (±2.5 psi)
|
|
Exhaust Pressure:
|
Atmospheric
|
|
Sample Flow Requirement:
|
<0.5 slpm
|
|
Inlet Gas Temperature:
|
0º to 100ºC (32º to 212ºF). Optimal results are obtained when the inlet gas temperature is maintained at 60ºC (140ºF)
|
|
Outputs:
|
Four line by 20 character LCD display
One self-powered 4 to 20 mA, into 100 to 500 ohm load analog output; can be configured for loop-powered operation
RS485 and RS232 serial ports
|
|
Alarms:
|
System alarm, concentration alarm, and data valid 30 VAC or 60 VDC max, 50 VA or 1A max, resistive
|
|
Environmental Conditions:
|
Ambient temperature range 10º to 30ºC (50º to 86ºF)
Optimal results are obtained when ambient is maintained within ±5ºC (±9ºF)
Relative humidity 90%, noncondensing
Pollution degree 2
Maximum altitude 2000 meters (6560 ft)
Installation category II
Indoor Use Only
|
|
Utility Requirements:
|
100 to 132 VAC or 230 VAC ±10%, 47 to 63 Hz, 185W
Instrument Air: 550 to 690 kPa (80 to 100 psi)
-40ºC dew point
|
|
Mounting Configuration:
|
19" rack
|
|
Dimensions(WxHxD):
|
48x17.7x50.9 cm (19x7x20 in)
|
|
Net Weight:
|
15.9 kg (35 lb)
|
|